Module Type |
9677 Module: 96wells x 0.2ml + 77wells x 0.5ml(standard) 96 Module: 96wells x 0.2ml(optional) Multi-functional Module:9677 Module + In situ loading plate(optional) 384 Module: 384wells(optional) |
9677 Module: 96wells x 0.2ml + 77wells x 0.5ml(standard) 96 Module: 96wells x 0.2ml(optional) Multi-functional Module:9677 Module + In situ loading plate(optional) |
Temperature Function |
The sample temperature range: 0℃∽ 100℃ Maximum Rising Speed: 5℃ /s Maximum Cooling Speed: 4℃ /s Sample Temp. Uniformity: ≤ ± 0.2℃ Sample Temp. Accuracy: ≤ ± 0.1℃ Temp. Control: Tube and sample table Variable temperature is adjustable |
The sample temperature range: 0℃∽ 100℃ Maximum Rising Speed: 5℃ /s Maximum Cooling Speed: 4℃ /s Sample Temp. Uniformity: ≤ ± 0.2℃ Sample Temp. Accuracy: ≤ ± 0.1℃ Temp. Control: Tube and sample table Variable temperature is adjustable |
Thermal Cover Function |
Thermal cover Temp.: 30℃∽ 112℃ (adjustable) Hot Acto. Shutdown Function: Yes |
Thermal cover Temp.: 30℃∽ 112℃ (adjustable) Hot Acto. Shutdown Function: Yes |
Programming Function |
The biggest step is 30, can do multiple nested loops; Time Increment/Decrement 1 ∽ 120s,can do Long PCR experiment; Temp. Increment/Decrement 0.1℃∽ 10.0℃, can do Touchdown PCR Experiment; Multiple User Password, ensure the user privacy and security program; Enter a base arrangement, automatic computing solution chain and annealing temperature, greatly improve the work efficiency; Program guide function, input a few necessary parameters can complete programming; Auto pause/power protection; Heat preservation, heat preservation time is infinite; |
Program is stored for more than 10000, can through the usb download unlimited program; The biggest step is 30, can do multiple nested loops; Time Increment/Decrement 1 ∽ 120s,can do Long PCR experiment; Temp. Increment/Decrement 0.1℃∽ 10.0℃, can do Touchdown PCR Experiment; Multiple User Password, ensure the user privacy and security program; Enter a base arrangement, automatic computing solution chain and annealing temperature, greatly improve the work efficiency; Program guide function, input a few necessary parameters can complete programming; Auto pause/power protection; Heat preservation, heat preservation time is infinite; |
Graded Function |
|
gradient uniformity ≤ ± 0.2℃; Temp. Accuracy: ≤ ± 0.1℃; Graded Range: 30℃∽ 99.9℃; Graded Temp. Range: ≤ 30℃; Temperature gradient distribution points 12 columns; |
Others |
Communication Interface: USB2.0, LAN; Voltage Range: Global 85V ∽ 265V, 50 ∽ 60Hz; Max Power: 600W; Size (L x W x H): 362x256x258mm; Net Weight: 7.4KG |
Communication Interface: USB2.0, LAN; Voltage Range: Global 85V ∽ 265V, 50 ∽ 60Hz; Max Power: 600W; Size (L x W x H): 362x256x258mm; Net Weight: 7.4KG |