Summary:
This instrument is designed and made as per standards DL/T706-99, IEC79-4:95. It is used to determine the lowest temperature of fire-resistant oil and other special oils when burning in a flask.
I. Main technical features
High precision AI temperature controller. The heating is uniform. Good accuracy. Three-point temperature detection: center of bottom, center on side wall and top.
II. Main technical specifications
1. Temperature range: Ambient to 800℃
2. Temperature control accuracy: Temperature difference ≤1℃ among the three points: center of bottom, center on side wall and top of flask.
3. Conical flask:200 ml
4. Ambient temperature: Room temperature~50℃
5. Relative humidity: <80%
6. Power supply: AC (220±10%) V, 50Hz
7. Maximum power consumption:2000W
8. Dimension:540mm×310mm×270mm
PT-SYD-706 Self Ignition Point Tester Accessories and Documents List
No. |
Name |
Unit |
Qty |
Memo |
1 |
PT-SYD-706 Self Ignition Point Tester |
Set |
1 |
|
2 |
Injector (250μL) |
Piece |
1 |
With needle |
3 |
Conical flask (200 ml) |
Piece |
1 |
Installed in machine |
4 |
Rectroreflector |
Piece |
1 |
Installed by user |
5 |
Power line (250V/6A) |
Piece |
1 |
|
6 |
Operation Manual |
Piece |
1 |
|
7 |
Quality Certificate |
Piece |
1 |
|
8 |
Repair Guarantee |
Piece |
1 |
|