Introduction:
	The instrument is designed and manufactured by ASTM D93 standard. The instrument is used to determine flash point of petroleum products in closed cup method. The petroleum product is heating in the airtight oil cup under the standard rated condition. The mixed gas formed by petroleum product steam with the surrounding air, when contact with the flame, the lowest temperature of the petroleum product, is the flash point of petroleum products in closed cup method.
	
	The instrument is composed of electric heating device, electric stirring device, electronic regulating device, ignition device, standard oil cup, etc. The instrument heat through controlled silicon pressure regulating control circuit. Start the motor, via driving flexible shaft, drive stirrer stir, oil temperature will be able to ensure uniformity and stability. 
	
	Technical Specifications
	1.Electric heating device	
	Electric wire 	800W
	Test oil heating rate	1 ~ 12 ℃ / min
	2. Electric mixing device	
	Constant speed motor 	105 rotate / min
	Driving Method: flexible shaft
	Agitating vane size: 8×40mm
	3. Mixing speed: 105 rotate / min
	Standard Oil Cup	
	Inner diameter: 50.8mm
	Depth: 56mm
	Tested oil capacity: about 70ml
	4. Ignition device:	
	Fire source: could use gas, liquefied gas and lighter gas
	Bore diameter	ø0.8mm
	5. Thermometer:	
	Internal scale or rod mercury thermometer
	Scale: -30 ℃ ~ 170 ℃ indexing 1 ℃
	Scale: 100 ℃ ~ 300 ℃ indexing 1 ℃
	6. Dimension: 300 × 250 × 200mm (length × width × height)
	7. Weight: about 5kg
	8. Relative humidity<85%
	9. Ambient temperature: 10 ~ 35 ℃
	10. Input power supply: AC 220V ± 10%