Main technical parameters:
Radio Frequency Generator (RF)
(1) Circuit type: Three-point self-oscillation circuit of capacitance
(2) Operating frequency: 40.68MHz ± 0.05%
(3) Frequency stability: <0.1%
(4) Output power :800-1200W
(5) Output power stability: ≤ 0 3%
(6) The intensity of electromagnetic field leakage: 30cm away from the generator
Electric field strength E: <2V / m
Sampling device
(1) Inner diameter of output working coil: 25mm, 3 turns
(2) Torch pipe: three concentric quartz torch pipes with outer diameter of 20mm
(3) Outer diameter of coaxial sprayer: 6mm
(4) Outer diameter of binocular spray chamber: 34mm
(5) specifications of argon flow meter:
① plasma gas flow meter: (100-1000) L / h, (1.6-16L/min)
② Auxiliary gas flow meter: (10-100) L / h, (0.16-1.66L/min)
③ Mass flow meter is used for carrier gas, and is controlled by computer (0-5L/min)
Optical Splitter
(1) Optical path: CzemyTurme
(2) Focal length: 1000mm
(3) Specifications of gratings: ion etching holographic grating
Groove density: 3600 lines / mm
Groove area: 80 × 110mm
(Optional groove density: 2400 lines / ram and groove area: 80 × 110mm)
(4) Reciprocal linear dispersion rate: 0. 26nm
(5) Resolution: ≤ 0.008nm
(6) Scanning wavelength range: 3600 lines / mm
Scanning wavelength range :195-500nm
2400 lines / mm
Scanning wavelength range :195-800nm
(7) Minimum step length of stepping motor driver: 0.0006nm
(8) Entrance and exit slit: 20um
(9) Size of reflecting mirror: (78 × 105 × 1 6) mm
(10) Lens: 1:1 imaging
(11) Constant-temperature device of optical splitter: 32 ℃ ± 1 ℃
Metering device
(1) PMT Specifications: R212UH or R928
(2) Negative high voltage of PMT: 200-1000V, stability <0.05%
(3) Current measurement range of PMT: 10-4 ~ 10-9 A
(4) V / F switching for signal acquisition: 1mV corresponds to 100Hz
Main technical characteristics:
1. New desktop no-op panel;
2. Using computer programs to control all the operating functions, including automatic ignition and so on;
3. Using mass flow meter for automatic control of carrier gas, which ensures the stability of sampling system;
4. All the operating functions are included in computer operating software, easy to use.
5. Low detection limit: ppb level.
6. High precision: RSD ≤ 2%.
7. High analysis speed: 10 or more elements per minute
8. Wide dynamic linear range (multi-element simultaneous measurement of 5-6 magnitude orders within linear range of working curves)
9. Humanized analysis software
List for Spare Parts
Name |
Quantity |
Name |
Quantity |
Manual |
1 |
Quarter bend |
2 |
Power line |
1 |
Fog cabinet |
2 |
Water Pipe 10 * 6.5 |
15 meter |
Fog chamber cover |
1 |
Gas Pipe4*6 |
15 meter |
Atomizer |
2 |
Gas pipe 4 * 6 |
2 meter |
Water-sealed bottle |
2 |
sewer pipe 6*8 |
1 meter |
Capillary |
2 |
wind-cone |
1 |
Quarter bend clamp |
1 |
Safety device 2A |
2 |
Fog cabinet seal ring |
2 |
ICP Analysis software |
1 |
Power strips |
1 |
Power line for host control |
1 |
|
|