This instrument is using the principle of ammonium molybdate spectrophotometric method, and it is ninth generation new products researched, developed, designed and manufactured according to the national standards: GB/T11893-1989.
Features
-Cold light source, narrow band interference, and the light resource life is 100,000 hours
-Large-size screen liquid crystal English display, which can directly read the concentration
-It can accurately test the content of total phosphorus in waste water
-The instrument can use national standard samples to calculate and store curve
-Among the built-in 100 curves, 5 standard curves do not need to adjust, can use directly, and other curves can be automatically repaired and stored
-It designs as modern popular mould and has transparent protecting cover which is antiseptic, acidproof, safe and beautiful
-It concludes three sets of timing system, which will automatically keep constant temperature and remind timely
-It can adjust and store temperature and time at will
Technical specifications
Measuring range |
0.001-20mg/L (subsection) |
Measuring error |
≦±5% |
repeatability |
±5% |
Measuring time |
35 ~ 50 minutes |
The number of curves |
100 pieces |
The stability of luminosity |
≦0.001A/10min |
Temperature indication error |
±0.5℃ |
Homogeneity of thermal field |
±0.5 |
The temperature of digestion |
122±0.5℃ |
The time of digestion |
30 minutes |
Indication error of digesting time |
±2s |
Environment temperature |
-5~40℃ |
The number of samples |
6/9/12/20/24/25/30/56 pieces/batches (to collocate according to the requirement of users) |
Working power supply |
AC220V±10%/50Hz |
Product size |
360x280x100mm |
Weight |
3Kg |