TA-1.0 Total Organic Carbon (TOC) Analyzer is the high-precision offline total organic carbon analysis instrument research and development by U-THERM INTERNATIONAL (H.K.) LIMITED independently. The product uses the electrical conductivity difference detection technology,adopting the principle of ultraviolet catalytic oxidation, with high accuracy and short response time. The instrument completely conforms to the requirements of the national pharmacopoeia, and it can meet the offline testing requirements for the deionized water such as pharmaceutical water (injection water and purified water) and ultrapure water.
Feature
1. There is no need of nitrogen, oxygen, acid reagent and oxidant, so the operation is simple and the cost is low;
2. The intelligent level is high, as it is equipped with automatic sampler, and the system adaptability validation can be done automatically;
3. It has the limit alarm design, and when the test sample exceeds the prescribed limit, the instrument makes automatic alarm;
4. It is in conformity with the test scheme as specified in the new edition of state pharmacopoeia;
Technical Parameters
Power Supply |
100-260VAC 50HZ,120W |
Conductivity Detection Range |
0.055卩s/cm-8.000卩s/cm |
Measure Range |
0.001mg/L 〜1.0mg/L(1 〜lOOOppb) |
Accuracy |
±3% |
RSD |
Repeatability < 3% |
Resolution |
0.001mg/L |
Response Time |
5mins |
Measuring Limit |
0.001 mg/L |
Sample Temperature |
1-99°C |
Applied Range |
Offline in the Laboratory、Cleaning Validation |
Audit trail |
Operational, storage > 3years |
Authority Management |
User name and password login , to meet FDA 21CFR PART 11 requirements |
Print Function |
External Printer |
Data Copy |
USB disk to export the data |
Display Screen |
Color Touch Screen |
Weight |
8.5KG |
Dimension(mm) |
400*240*270 |