Welcome: U-THERM INTERNATIONAL (H.K.) LIMITED

   Home   |   Product   |   Split Spraying Laser Particle Size Analyzers

Particle Size Analyzers

  • Split Spraying Laser Particle Size Analyzers
Split Spraying Laser Particle Size Analyzers

Split Spraying Laser Particle Size Analyzers

  • Model: PSA318C
  • Sample introduction : Open
  • Spray area range : 1-10m
  • Lens protection : Double air curtains
  • Product description: Split Spraying Laser Particle Size Analyzers
  • INQUIRY

PSA318C is a split-style laser particle size analyzer dedicated to studying sprays. It can also study particles dispersed within air or liquids, or emulsions if it’s used in combination with dry/wet dispersing systems.

 

Main Features:

1. Measurement principle: It carries out non-contact measurements on sprays and the calculation is based on Fraunhofer light diffraction theory. An unconstrained fitting program processes the data and can obtain accurate results for any sorts of size distributions.

2. Novel technique: The spectrum amplification technique extended the measurable range up to 2000um, without the need to increase the instrument dimension. Multiple detector arrays measure the scattered laser at a wide range of angles and this ensures reliable results for both big and small particles.

3. Split-style With such a design the spray range can be adjust from 1 to 10 meters. This meets the needs of many users, and the measurements have non-contact and non-disturbance to the sprays. Such an open design makes It to be able to study particles dispersed within air or liquids, or emulsions if it’s used in combination with the dry/wet dispersing systems

4. Contamination proof The instrument is protected from contamination of the sprays.

5. Rich software functions: Beside the basic function of analyzing the size distributions, the software also has the following features:

a) The output has rich content including: size distribution, R-R distribution, logarithmic distribution, statistics within different size ranges, volume distribution, number distribution, the percentage of particles of a certain size, or a certain size range, and comparison of multiple results.

b) The output could be in Word, Excel, Bmp pictures and Text formats, and this makes easy viewing and referencing the results at different situations.

c) It supports English and Chinese languages, and it can be changed to many other language versions according to the user’s needs.

d) Calibration: Throughout the measurable size ranges the instrument is calibrated with China’s standard reference particulate materials. The resolution and accuracy are guaranteed.

 

Specifications

Model

PSA318A

PSA318B

PSA318C

Execution standard

ISO13320GB/T19077.1-2008Q/JW001-2006

Structure

Split-type

Measurement range

4.6-323μm

15-711μm

4.6-2000μm

Number of channel

30

30

60

Accuracy error

<3%standard reference materials

Reproducibility error

<3%standard reference material

Laser

LD pump laser; λ=532nmp>40mw

Sample introduction

Open

Spray area range

1-10m

Lens protection

Double air curtains

Adjusting range

Horizontal moving

±20mm

Lifting

±20mm

rotation

±20°

Dimension

500mm×254mm×520mm×2

Weight

45Kg+23Kg

 

 

Shipping Package

Name

Quantity

PSA318

1

Laser particle analyzer software

including Encryption dog

1

Low noise fan

2

Manual

1

Standard Reference material

1 bottle

Gears

1 set

Power line

2

Signal cable

2

USB conversion line

2

 

Applications:

Studying sprays appearing in the study of fire fighting, pesticides, air engine, spray granulation, nozzles and aerosol sprays. 