Principle:
	The rapid plasticity value P0 is rapidly compressed by the main body to a cylindrical sample between two parallel compacts to a fixed thickness of 1 mm, and the test sample is kept under compression for 15 s to achieve a temperature balance with the parallel plate, and then tested. A constant pressure of 100 N ± 1 N was applied and held for 15 s. At the end of this phase, the measured thickness of the test is accurately measured by the observation instrument as a measure of plasticity. 
		  
		Technical parameters: 
						Host 
						Temperature
  control 
						100±0.5℃ 
						Load
  value 
						100±1N 
						Power
  supply 
						AC
  220V±10% 
						Upper
  plate size 
						φ7.3±0.02mm,
  φ10±0.02mm, φ14±0.02mm 
						Upper
  plate effective height 
						3.2mm 
						Lower
  plate 
						φ15mm 
						Aging
  Chamber 
						Temperature
  control 
						140±0.2℃ 
						Power
  supply 
						AC
  220V±10% 
						Timing
  time 
						30min
  ± 5s 
		  
		 
	 
		 
	
The plastic retention rate PRI was measured by aging the sample at a temperature of 140 ° for 30 min through a high-precision special aging box, and the plastic value P30 was measured, and the PRI value was calculated by calculation.
The PRI value is a measure of the thermal oxidation of natural rubber, and a higher value indicates a better thermal oxidation resistance.
Features:
1. Direct electronic temperature control, fast heating and high precision
2. Direct loading with standard weights overcomes the shortcomings of instability and zero drift caused by the stress of the loaded electromagnetic spring.
Standards: 
ISO2007,  ISO2930
Composition:
It consists of HOST, punching machine (including cutting knife), aging test chamber and thickness gauge.