Features
◆ 4 calibration points can be chosen, the meter provides calibration guide and automatic checking function.
◆ The meter is able to switch among conductivity, TDS, salinity and resistivity measurement mode. For TDS measurement, the coefficient is adjustable.
◆ The meter is able to recognize up to 8 types of conductivity standard solutions. There are two options of standard buffer solutions: USA series and Chinese series (CH), there is also customer-defined solution calibration.
◆ The meter has an automatic range feature with automatic non-linear temperature compensation for pure water and high purity water with conductivity values lower than 10uS/cm greatly improving accuracy and suitable for the electricity, microelectronics and pharmaceutical industries.
◆ Equipped with calibration solution and portable case and suited in field and on site.
◆ Meter meets IP57 rating.
Specifications
Conductivity |
Range |
Cond:0~200mS/cm divided into five ranges: |
Resolution |
0.01/0.1/1μS/cm 0.01/0.1 mS/cm |
|
Accuracy |
±1.0% FS ±1digit |
|
Temperature compensation |
(0 ~ 50) °C (manual or automatic) |
|
Electrode constant |
0.1 / 1 / 10 cm-1 |
|
Temperature |
Range |
0~100°C |
Resolution |
0.1℃ |
|
Accuracy |
±0.5℃±1digit |
|
Other Technical Parameters |
Data storage |
500 groups |
Storage content |
Serial number, date, time, measuring value, measuring unit and temperature value |
|
Power |
AAA batteries × 3 (1.5V× 3) / DC5V adaptors (USB port) |
|
Dimension & Weight |
Meter: (88×170×33 )mm / 313g;Portable case: (360×270×76) mm/1.6kg |
|
IP rating |
IP57 dustproof and waterproof |