Welcome: U-THERM INTERNATIONAL (H.K.) LIMITED

   Home   |   Product   |   Cold Filter Plugging Point Filter

Low Temperature Tester

  • Cold Filter Plugging Point Filter
Cold Filter Plugging Point Filter

Cold Filter Plugging Point Filter

  • Model: PT-LC-2
  • t adopts integral structure
  • It adopts light-touch and digital display technology
  • The data set is easy and the timing shows strikingly
  • Product description: Cold Filter Plugging Point Filter
  • INQUIRY

Summary

The instrument is designed and made as per the Industry Standard of People’s Republic of China SH/T 0248-2006 Standard Diesel and Domestic Heating Fuels - Determination of Cold Filter Plugging Point. Being equipped with refrigeration bath and other accessories, it can be used to determine cold filter plugging point of distillate fuels, including fuels containing fluidity improver or other additive, and used for diesel engines and domestic heating devices.

I. Main technical features
1. It adopts integral structure. Easy to use and good-looing.
2. It adopts light-touch and digital display technology. The data set is easy and the timing shows strikingly. It stops automatically after the timing is over.
3. It equips a flowmeter to control and show the flow rate of air suction.
4. The accessories are fully equipped. It equips two vacuum bottles to control the air suction. It can meet standard SH/T 0248-2006.
5. The instrument can be singly used to do CFPP determination equipping with suitable cooling bath. It can also be used to do CFPP determination equipping with suitable low temperature tester.

II. Main technical specifications
1. Power supply:  AC 220V±10%, 50 Hz
2. Pumping pressure:  1961 Pa (200 mm H2O)
3. Test cup:  Glass, flat bottom, circular type; inner diameter
31.032.0 mm; Thickness 1.01.5mm; Height 115125mm
There is a mark line at 45 ml of the cup.
4. Filter assembly:  It is made of cooper; Size of hole on the filter sieve is 45 μm (330 meshes)
5. Working environment: (1) Ambient temperature: -10+35℃ (2) Relative humidity: ≤85%
6. Maximum power consumption:  150 W
7. Dimension:  250mm×150mm×380mm (Vacuum bottles are not included)

 

PT-LC-2 Cold Filter Plugging Point Filter Accessories and Documents List 

No.

Name

Unit

Qty

Memo

1

PT-LC-2 Cold Filter Plugging Point Filter

Set

1

 

2

Test cup

Piece

1

 

3

Cold filter tube plug

Piece

1

 

4

Pipette

Piece

1

 

5

Filter assembly

Piece

1

Copper

6

Sieve

Piece

2

One is already installed on the filter assembly

7

Syringe(5ml)

Piece

1

 

8

Silicone pipe (Φ6, thickness 1.5)

Length430,320,140

Piece

Each 1

 

9

Silicone pipe (Φ5, thickness 1.0)

Length800

Piece

1

 

10

Silicone pipe (Φ2, thickness 0.75)

Meter

0.75

Match up with the syringe

11

Fuse 2A (Ф5×20)

Piece

2

 

12

Thermometer -80+20, scale division 1

Piece

1

 

Thermometer -38+50, scale division 1

Piece

1

 

13

Sealing grease

Bottle

1

Used to seal the bottle stopper

14

Sealing tape

Box

1

sed to seal glass pipes

15

Operation manual

Piece

1

 

16

Quality Certificate

Piece

1

 

17

Repair Guarantee

Piece

1