Welcome: U-THERM INTERNATIONAL (H.K.) LIMITED

   Home   |   Product   |   Automatic Vapor Pressure Tester (Reid Method)

Vapor Pressure Tester

  • Automatic Vapor Pressure Tester (Reid Method)
Automatic Vapor Pressure Tester (Reid Method)

Automatic Vapor Pressure Tester (Reid Method)

  • Model: SYD-8017A
  • Power supply: AC(220±10%)V, 50Hz
  • Maximum power consumption: ≤1700W
  • Test bomb: Can do 3 bomb tests at the same time
  • Product description: Automatic Vapor Pressure Tester (Reid Method)
  • INQUIRY

Introduction:

This instrument is made and designed as per national standard of People’s Republic of China GB/T 8017 Petroleum Products- Determination of vapor pressure- Reid method. It can also be used to do determination according to ASTM D323 Standard Test Method for Vapor Pressure of Petroleum Products (Reid Method). It is used to make determination for vapor pressure of gasoline, volatile crude oil and other volatile petroleum products. It cannot be used to determine the vapor pressure of LGP. It is fully-automatic vapor pressure tester by PC control and operation.

 

I. Main technical features

1.The instrument is made of the upper machine (PC) and the lower machine ( the host instrument ),they are connected with USB communication port ,it can do the operation on the computer.

2.The computer can determine, calculate and control to system parts. It has advantages of high automation, accurate test results, good repeatability and easy operation.

3. Advanced heating technology do separation of water and electricity meanwhile lower the height of water in bath. No blind angle in bath. It can save water and energy and also convenient to be cleaned.

4. The motion mode of the bomb can change from the horizontal axial to traditional 350-degree rotation into a "seesaw" movement, which can make the sample stirring fully and it can meet the requirements of standard, the results of the vapor pressure is more accurate.

 

II. Main technical specifications

1. Power supply: AC(220±10%)V, 50Hz

2. Maximum power consumption: ≤1700W

3. Test bomb: Can do 3 bomb tests at the same time.

4. Water bath temperature: 37.8℃

5. Temperature control accuracy: ±0.1℃

6. Heating power: 1600W

7. Pressure range: (0200)kPa or (029)psi

8. Ambient temperature: (535)

9.Relative humidity: ≤85%

10.Dimension: 600mm×500mm×512mm

Note: Equipped with PC, a pressure calibration device is enclosed

 

PT-SYD-8017A Automatic Vapor Pressure Tester (Reid Method) Accessories and Documents List 

No.

Name

Unit 

Qty 

Memo 

1

PT-SYD-8017A Automatic Vapor Pressure Tester (Reid Method)

Set

1

 

2

Laptop

Set

1

 

3

Oxygen bomb test assembly

Set

3

 

4

Thermometer 34~42, sale division is 0.1

Piece

1

Specially used in saturation vapor pressure

5

Professional usage of saturation vapor pressure

Piece

1

 

6

Drain pipe (Φ18 × Φ13)

Piece

1

 

7

Fuse 15A (Φ5×20)

Piece

2

 

8

USB communication cable

Piece

1

 

9

Installation CD

Piece

1

 

10

Calibrator of sturation vapor pressure

Set

1

 

11

Air chamber sealing plug

Piece

3

 

12

Liquid chamber sealing plug

Piece

3

 

13

Operation manual

Piece

1

 

14

Quality certificate

Piece

1

 

15

Product Warranty Card

Piece

1

 

 

PT-SYD-8017A Automatic Vapor Pressure Tester (Reid Method)'s Reagent and Auxiliary Equipment

No. 

Name 

Specification 

Quantity 

Memo 

1

Naphtha

500ml

1 bottle 

As the cleaning agent to for oxygen compound gasoline sample

2

Acetone

500ml

1 bottle

3

Hydrargyrum

500g

1 bottle

To calibrate the device

4

Refrigerator

Smallest

1 set

Constant temperature 0-1  


Notes
1.The reagent and equipment for installing the instrument are supplied by end-users themselves.
2.The usage, management and saving of poisonous and harmful reagent, pls operate according to the relative principle.
3.PLS read the operation manual and industry standard before using the instrument. 

PRODUCT CATEGORIES

LATEST NEWS