Overview
The instrument is designed and made as per industrial standard of People’s Republic of China SH/T 0193 Lubricating oils-Determination of oxidation stability-Rotating pressure vessel method and ASTM D2272 Standard test Method for Oxidation Stability of Steam Turbine Oils by Rotating Pressure Vessel.
The instrument is used to determine the oxidation stability of steam turbine with the same composition (oil base oil and additive) .Also can be used to determine new mineral insulating oil containing 2, 6-BHT.
I. Main technical features
1. This instrument is fully designed as per standard SH/T0193. It can automatically determine pressure, record data, calculate oxidation time and finish the determination.
2. The test procedures can be done automatically. No need for human interference. The test precision is high.
3. The test curves and results will be shown on the screen and can be restored and printed if needed.
II. Main technical specifications
1. Power supply: AC220V±10%
2. Heating tube power: 2500W
3. Range for pressure sensor: (0~1.6)MPa, accuracy:±2‰
4. Temperature control range for oil bath:Room temperature~200.0℃, continuously adjustable.150.0℃ in common use.
5. Temperature control accuracy: ±0.1℃
6. Test sample:Two-bomb design, can do two samples at one time. Convenient to do parallel test.
7. Rotation speed: (100±5)r/min
8. Included angle between oxygen bomb and water level: 30°
9. Volume of oil bath: 30L
10.Dimension: 550mm×800mm×1000mm
11.Net weight: About 45Kg
PT-SYD-0193 Automatic Lubricating Oils Oxidation Stability Tester Accessories and Documents List
No.
Name
Unit
Qty
Memo
1
PT-SYD-0193 Automatic Lubricating Oils Oxidation Stability Tester
Set
1
2
Notebook computer
Set
1
3
Oxygen bomb assembly
Set
2
4
Glass sample cup
Piece
4
5
PTFE cover
Piece
2
6
Oxygen bomb sealing ring
Piece
4
7
Test copper wire (diameter 1.63mm, made of Φ47 × 21 ring ring copper ring 2)
Piece
6
8
Oxygen bomb wrench
Piece
1
9
Connecting line and joint of oxygen cylinder
Piece
1
10
USB connection
Piece
1
11
Rod type mercury thermometer 130 ℃ ~ 150 ℃, scale division 0.1 ℃
Piece
1
12
Fixed spring leaf (on the PTFE cover)
Piece
2
13
Winding machine
Set
1
14
Fuse
(20A/500V Φ10×38)
Piece
2
15
Operation manual
Piece
1
16
Quality certificate
Piece
1
17
Product warranty card
Piece
1
PT-SYD-0193 Automatic Lubricating Oils Oxidation Stability Tester's Reagent and Auxiliary Equipment
No. |
Name |
Specification |
Quantity |
Memo |
1 |
Silicone oil |
Flash point above 200℃ |
30L |
0193B no need preparation |
2 |
Miscella |
2# or 3# 500ml |
1 bottle |
for cleaning |
3 |
Acetone |
Analytical pure 500ml |
1 bottle |
|
4 |
Isopropanol |
Analytical pure 500ml |
1 bottle |
|
5 |
Normal heptane |
Analytical pure 500ml |
1 bottle |
|
6 |
Potassium hydroxide |
Analytical pure 500g |
1 bottle |
|
7 |
Distilled water |
2 grades |
Less |
|
8 |
Silicon carbide paper |
Particle size 100# |
5 pieces |
|
9 |
Oxygen |
Purity 99.5% |
1 bottle |
With reducing valve |
10 |
Silicone grease |
|
1 box |
|
Notes:
1.The reagent and equipment for installing the instrument are supplied by end-users themselves.
2.The usage, management and saving of poisonous and harmful reagent, pls operate according to the relative principle.
3.PLS read the operation manual and industry standard before using the instrument.